Mindray Bc 2800 Vet User Manual [verified]
| Error Code | Message | Probable Cause | Manual Solution | | :--- | :--- | :--- | :--- | | E001 | Sample Probe Blockage | Clotted blood or fibrin | Run “Probe Clean” cycle. Manually probe with wire tool. | | E010 | Reagent Empty | Diluent or Lyse bottle empty | Replace reagent. Prime the reagent path (Maintenance > Prime). | | E030 | Flow Cell Dirty | Protein buildup | Run 3 cycles of “Flow Cell Clean.” If persists, soak in 10% bleach. | | E100 | HGB Zero Failed | Dirty HGB cell or LED aging | Clean the HGB flow cell. If LED fails, replace (service call). | | WBC Flag | “Turbidity/HB Interf?” | Lipemic or hemolyzed sample | Centrifuge sample and replace plasma with saline. Re-run. |
Six counting modes including whole blood and prediluted options. mindray bc 2800 vet user manual
: Stores up to 20,000 results, including histograms, for easy patient history review. Pro-Tips for Daily Operation | Error Code | Message | Probable Cause
The BC-2800 Vet strictly requires EDTA-K2 (purple top) tubes. Using the wrong anticoagulant (like Heparin) will result in erroneous platelet clumping and WBC distortions. Prime the reagent path (Maintenance > Prime)
Whether you are troubleshooting a stubborn “Flow Cell Dirty” error or teaching a new technician the proper shutdown procedure, keep a copy of the manual—physical or PDF—within arm's reach of the analyzer. When in doubt, consult the manual before pressing any button.